Dose and Image Quality in Spectral Computed Tomography Compared to Helical Protocols
DOI:
https://doi.org/10.53903/01212095.402Keywords:
Radiation dose, Computed tomography, Dual-energy CT, Diagnostic reference levelsAbstract
Objectives: Dual-energy computed tomography (DECT) provides additional diagnostic information compared with conventional single-energy CT (SECT), particularly in oncology. This study aimed to calculate the typical dose values for DECT protocols and compare them with those of SECT, as well as to evaluate the image quality of both modalities. Methods: A retrospective analysis of DECT studies was conducted at an oncology institution. Typical radiation dose values (CTDIvol and DLP) were determined according to the ICRP 135 method. Image quality was assessed in a subset of abdominal studies by three radiologists using a predefined ordinal scale to evaluate noise, sharpness, and overall contrast. Results: The typical dose values for DECT were lower than those obtained with equivalent SECT protocols (CTDIvol: 7.69–9.37 mGy; DLP: 242–480 mGy·cm). Qualitative assessment showed that 95 % of the studies had minimal or mild noise, 98 % were rated as acceptable or higher in sharpness, and 96 % were rated as acceptable or higher in overall contrast. Conclusions: In our experience, DECT did not result in an increased radiation dose compared with SECT and provided additional diagnostic information with adequate image quality. These findings support the use of DECT as a clinically feasible alternative that balances radiation dose and image quality.
Downloads
References
Kong D, Chen X, Gao P, Zhao K, Zheng C, Zhou H. Diagnostic accuracy of contrastenhanced dual-energy computed tomography for detecting metastatic lymph nodes in patients with malignant tumors: a systematic review and meta-analysis. Quant Imaging Med Surg. 2023;13(5):3050-65.
Wang T, Fan Z, Zou L, Hou Y. Can quantitative parameters of spectral computed tomography predict lymphatic metastasis in lung cancer? a systematic review and meta-analysis. Radiother Oncol. 2023;183:109643. https://doi.org/10.1016/j.radonc.2023.109643.
Chen M, Jiang Y, Zhou X, Wu D, Xie Q. Dual-energy computed tomography in detecting and predicting lymph node metastasis in malignant tumor patients: a comprehensive review. Diagnostics (Basel). 2024;14(4):377. https://doi.org/10.3390/diagnostics14040377.
Vañó E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, et al. ICRP Publication 135: Diagnostic reference levels in medical imaging. Ann ICRP [Internet]. 2017;46(1):1-144. Disponible en: http://journals.sagepub.com/doi/10.1177/0146645317717209
Kanal KM, Butler PF, Sengupta D, Bhargavan-Chatfield M, Coombs LP, Morin RL. U.S. diagnostic reference levels and achievable doses for 10 adult CT examinations. Radiology. 2017;284(1):120-33.
Cadavid L, Karout L, Kalra MK, Morgado F, Londoño MA, Pérez L, et al. Setting up regional diagnostic reference levels for pediatric computed tomography in Latin America: preliminary results, challenges and the work ahead. Pediatr Radiol. 2024;54(3):457-67.
Oliveira Bernardo M, Karout L, Morgado F, Ebrahimian S, Sarmet Santos A, Amorim C, et al. Establishing national clinical diagnostic reference levels and achievable doses for CT examinations in Brazil: A prospective study. Eur J Radiol. 2023:169:111191. https://doi.org/10.1016/j.ejrad.2023.111191.
McCollough CH, Leng S, Yu L, Fletcher JG. Dual- and Multi-Energy CT: principles, technical approaches, and clinical applications. Radiology. 2015;276(3):637-53.
Uhrig M, Simons D, Kachelrieß M, Pisana F, Kuchenbecker S, Schlemmer HP. Advanced abdominal imaging with dual energy CT is feasible without increasing radiation dose. Cancer Imaging. 2016;16(1):15.
Schmidt D, Söderberg M, Nilsson M, Lindvall H, Christoffersen C, Leander P. Evaluation of image quality and radiation dose of abdominal dual-energy CT. Acta Radiol. 2018;59(7):845-52.
Wortman JR, Shyu JY, Dileo J, Uyeda JW, Sodickson AD. Dual-energy CT for routine imaging of the abdomen and pelvis: radiation dose and image quality. Emerg Radiol. 2020;27(1):45-50.
Lenga L, Leithner D, Peterke JL, Albrecht MH, Gudauskas T, D’Angelo T, et al. Comparison of radiation dose and image quality of contrast-enhanced dual-source CT of the chest: single-versus dual-energy and second-versus third-generation technology. AJR Am J Roentgenol. 2019;212(4):741-7.
Ahmad M, Jacobsen MC, Thomas MA, Chen HS, Layman RR, Jones AK. A Benchmark for automatic noise measurement in clinical computed tomography. Med Phys. 2021;48(2):640-7.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Revista Colombiana de Radiología

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.
La Revista Colombiana de Radiología es de acceso abierto y todos sus artículos se encuentran libre y completamente disponibles en línea para todo público sin costo alguno.
Los derechos patrimoniales de autor de los textos y de las imágenes del artículo como han sido transferidos pertenecen a la Asociación Colombiana de Radiología (ACR). Por tanto para su reproducción es necesario solicitar permisos y se debe hacer referencia al artículo de la Revista Colombiana de Radiología en las presentaciones o artículos nuevos donde se incluyan.
| Article metrics | |
|---|---|
| Abstract views | 611 |
| Galley views | 252 |
| PDF Views | 252 |
| HTML views | 0 |
| Other views | 0 |